Syllabus · Ultrasound and Doppler physics
Ultrasound elastography
DNB Radiodiagnosis notes · a free sample topic by Dr. Anoop Phatnani, DNB Radiodiagnosis
Answer in a page
The core is always the same three blocks (principle, techniques, applications); what changes is which organ is asked in depth.
| Question shape | Write |
|---|---|
| "Principle, techniques and clinical applications" (3+3+4, the commonest) | Definition, Young's modulus, the two diagrams; classification table of techniques; applications by organ with one line each |
| "Principle" alone (5 marks) | Definition, stress and strain, the two ways of measuring stiffness, E = 3ρc², units, the diagram |
| "Applications in diffuse liver disease" or "role in liver disease" (3-4 marks) | Technique of measurement, fibrosis staging, cut-offs with their source, portal hypertension, confounders |
| "Shear wave elastography" short note | Principle, three types, units, applications, limitations |
| "Clinical applications of compression elastography in detail" | Organ by organ for strain imaging: breast, thyroid, cervix, prostate, nodes, musculoskeletal, endoscopic |
| "Role in breast pathologies" (10 marks) | Technique, Tsukuba score, strain ratio, E/B ratio, shear wave thresholds, use with BI-RADS, pitfalls |
| "Role of thyroid elastography" (5 marks) | Scores, strain ratio, shear wave values, where it helps, limitations |
| "Sonoelastography in obstetrics and gynaecology" (5 marks) | Cervix, placenta, fibroid versus adenomyosis, endometrium, ovary, safety |
| "Ultrasound versus MR elastography for liver fibrosis" | Comparison table |
Always state that cut-offs differ between vendors and techniques, and name the guideline you are quoting.
Principle
Definition. Elastography is imaging of the mechanical stiffness of tissue. It is the imaging equivalent of palpation. Disease changes stiffness: fibrosis, most cancers and inflammation make tissue harder. When ultrasound is used to do it, it is called ultrasound elastography or sonoelastography.
Why B-mode cannot do this. B-mode depends on the bulk modulus, which decides the speed of the ordinary (longitudinal) sound wave. It varies very little across soft tissues, which is why the scanner can assume 1540 m/s everywhere. The shear modulus, which decides how easily tissue deforms, varies many-fold between tissues and between normal and diseased tissue. Elastography images that property.
The physics
- Stress (σ): force per unit area applied to the tissue. Unit: pascal (Pa = N/m²).
- Strain (ε): the deformation produced, as change in length divided by original length. No unit.
- Young's modulus (E) = stress / strain. A stiff tissue needs a large stress to produce a small strain, so E is high. Unit: kilopascal (kPa).
- Shear modulus (G): resistance to a shearing (sideways) force. In soft tissue, which is almost incompressible, E = 3G.
- A shear wave is a transverse wave: particles move at right angles to the direction of travel. It is slow (about 1-10 m/s in soft tissue, against 1540 m/s for the ultrasound beam), it is attenuated quickly, and it does not travel through simple (non-viscous) fluid.
- Shear wave speed (c) depends on stiffness: G = ρc², so E = 3ρc², where ρ is tissue density (taken as 1000 kg/m³). With c in m/s, E in kPa is simply 3 × c². Example: 1.5 m/s is about 6.8 kPa.
Every elastography method has three steps
- Excite: apply a force to deform the tissue (hand pressure, physiological pulsation, a mechanical piston, or a focused ultrasound push pulse).
- Track: follow the tissue response with ordinary ultrasound pulses (speckle tracking by correlation of radiofrequency echoes before and after).
- Display: show the result as a colour map or greyscale map (elastogram) or as a number.
Two things can be measured, and this gives the two families
| Strain imaging | Shear wave imaging | |
|---|---|---|
| What is measured | How much the tissue deforms along the beam | How fast a shear wave travels across the beam |
| Stress | Unknown, so E cannot be calculated | Not needed; speed alone gives E |
| Result | Qualitative or semi-quantitative (colour, score, ratio) | Quantitative (m/s or kPa) |
Acoustic radiation force impulse (ARFI). A focused ultrasound pulse, much longer than an imaging pulse (a fraction of a millisecond), transfers momentum to tissue as it is absorbed and pushes the tissue at the focus by a few micrometres. The scanner can either measure that displacement on the beam axis (a form of strain imaging) or track the shear waves that spread sideways from the push (shear wave elastography). The same transducer pushes and tracks.
Assumptions behind the number: tissue is taken to be homogeneous, isotropic, incompressible and purely elastic. Real tissue is viscoelastic, so shear wave speed changes with the frequency of the wave. This is one reason why different machines give different values in the same patient.
Draw this
For a 3-mark "principle" part, panels A and B with their formulae are enough.
Techniques
| Technique | Excitation | Measures | Output | Image guidance | Example |
|---|---|---|---|---|---|
| Strain (compression) elastography | Manual probe pressure, or cardiac, vascular or respiratory motion | Axial strain | Colour or grey map, score, strain ratio | Yes, real time | Breast, thyroid, cervix |
| ARFI strain imaging | Ultrasound push pulse | Displacement on the beam axis | Qualitative map | Yes | Siemens Virtual Touch Imaging |
| Transient elastography (TE) | Mechanical piston on the skin, 50 Hz | Shear wave speed along one line | One number, kPa | No (A-mode and M-mode only) | FibroScan |
| Point shear wave elastography (pSWE) | ARFI push at one focus | Mean shear wave speed in one small fixed box | One number, m/s or kPa | Yes, B-mode | Siemens VTQ, Philips ElastPQ |
| 2D shear wave elastography (2D-SWE) | ARFI pushes at many points | Shear wave speed pixel by pixel | Colour map plus numbers, m/s or kPa | Yes, real time or near real time | SuperSonic Imagine (Aixplorer), GE, Canon, Philips ElastQ |
Transient elastography, pSWE and 2D-SWE are all shear wave methods. Many texts use "shear wave elastography" for the two ARFI-based methods only and keep "transient elastography" as a separate name; some use the terms interchangeably. Write the classification as above and you are covered either way.
Strain (compression) elastography
Also called static, quasi-static or real-time elastography.
How it is done
- A high-frequency linear probe is held perpendicular to the skin. The lesion is placed in the centre of a colour box that also includes plenty of normal surrounding tissue (the lesion should fill no more than about a quarter to a half of the box).
- The operator applies light, repeated compression and release, or simply holds the probe still and lets breathing and pulsation move the tissue. Most machines show a bar or a graph that confirms the compression is adequate.
- The machine compares radiofrequency echoes before and after compression, calculates the displacement at each depth, and from its gradient the strain.
- The elastogram is shown beside or over the B-mode image. On most systems blue is hard, green intermediate and red soft; some vendors use the opposite, so always read the colour bar. On a greyscale map, dark is hard.
How it is read
| Method | What it is |
|---|---|
| Colour pattern score | Visual score of the colours in and around the lesion: Tsukuba (breast), Rago or Asteria (thyroid) |
| Strain ratio | Strain of reference tissue (fat in the breast, normal parenchyma in the thyroid) divided by strain of the lesion, both at about the same depth. A higher ratio means a stiffer lesion |
| E/B ratio (size ratio) | Lesion length on the elastogram divided by its length on B-mode. Cancers look larger on the elastogram because of the stiff desmoplastic reaction around them |
| Fat-to-lesion ratio, strain histogram, hard-area ratio | Vendor-specific variants of the above |
Strengths: available on most machines, cheap, real time, high spatial resolution, no extra acoustic output beyond B-mode.
Weaknesses
- Relative only: it shows contrast between a lesion and its surroundings, so it is poor for diffuse disease where everything is equally stiff.
- Operator dependent: too much pre-compression makes all tissue look stiff (stiffness rises non-linearly with pressure).
- Stress falls with depth, so it works to about 3-4 cm only.
- The lesion can slip sideways out of the plane during compression.
- Ratios and scores are not comparable across vendors.
Shear wave elastography
Principle. A force is applied briefly and the shear waves it produces are tracked. Their speed is measured and reported in m/s, or converted to Young's modulus in kPa with E = 3ρc². Because the result does not depend on how hard the operator presses, it is quantitative and more reproducible than strain imaging.
Transient elastography (vibration-controlled transient elastography, FibroScan)
- The first shear wave method, and the one with the most validation in liver disease. It is not an imaging ultrasound machine: there is no B-mode picture.
- The probe carries a single-element transducer on a vibrating piston. The piston taps the skin at 50 Hz; the transducer tracks the resulting shear wave along its own axis.
- M probe: 3.5 MHz, measures at 25-65 mm from the skin. XL probe: 2.5 MHz, 35-75 mm, for obese patients. An S probe exists for children.
- It samples a cylinder of liver about 1 cm wide and 4 cm long, roughly 100 times the volume of a biopsy core.
- The patient lies supine with the right arm raised; the probe is placed in a right intercostal space over the right lobe. Ten valid readings are taken and the median is reported in kPa (the machine reads up to 75 kPa).
- Reliable when the interquartile range divided by the median (IQR/M) is 30% or less.
- The same probe gives the controlled attenuation parameter (CAP) in dB/m (100-400), an estimate of steatosis.
- It fails with ascites (the mechanical wave does not cross fluid), in obesity and with narrow intercostal spaces.
Point shear wave elastography
- Done on a conventional scanner with a curved probe. The operator sees the B-mode image and places a small fixed box (about 1 cm) in liver free of vessels and ducts.
- One ARFI push is delivered; shear wave speed is averaged over the box and shown as a single value.
- Works through ascites, because the push is generated inside the liver.
- Ten readings, median and IQR/M, as for transient elastography.
2D shear wave elastography
- ARFI pushes are delivered at many points. In the SuperSonic method they are focused at successively deeper points faster than the shear wave travels, producing a cone-shaped ("Mach cone") wavefront; ultrafast plane-wave imaging at thousands of frames per second then tracks it.
- A colour map of stiffness is laid over the B-mode image. One or more regions of interest are drawn in it, giving mean, maximum, minimum and standard deviation.
- Many machines add a quality or propagation map (parallel, evenly spaced wavefront lines mean a good acquisition) and a stability index.
- A 3D version exists for the breast.
Clinical significance in one line each: staging liver fibrosis without biopsy and predicting portal hypertension (the main use); adding specificity to breast and thyroid ultrasound; targeting prostate biopsy; assessing tendon, muscle, lymph node, kidney graft and spleen.
Limits: depth (ARFI methods are unreliable beyond about 6-7 cm), motion, obesity, vendor-specific values, and no signal in fluid.
Clinical applications
| Organ | Main use | Usual technique |
|---|---|---|
| Liver | Staging fibrosis, ruling compensated advanced chronic liver disease in or out, portal hypertension | TE, pSWE, 2D-SWE |
| Spleen | Predicting clinically significant portal hypertension and varices | TE, pSWE, 2D-SWE |
| Breast | Benign versus malignant; refining BI-RADS 3 and 4a | Strain, 2D-SWE |
| Thyroid | Selecting nodules for fine-needle aspiration; thyroiditis | Strain, SWE |
| Prostate | Targeting biopsy to stiff peripheral-zone foci (transrectal); about 35 kPa reported as a 2D-SWE cut-off. False positives: hyperplasia, calcification, prostatitis | Strain, 2D-SWE |
| Lymph nodes | Metastatic nodes are stiffer; choosing the node or the part of the node to sample | Strain, SWE |
| Musculoskeletal | Tendinopathy (a diseased tendon softens), muscle injury and spasticity, plantar fascia, soft tissue masses | Strain, SWE |
| Kidney | Fibrosis in chronic kidney disease and in the transplant kidney (results inconsistent because the kidney is anisotropic) | SWE |
| Pancreas and gut | Endoscopic ultrasound elastography of pancreatic masses and nodes; fibrotic versus inflammatory stricture in Crohn disease | Strain, SWE |
| Obstetrics and gynaecology | Cervix, fibroid, adenomyosis, endometrium, placenta | Strain, SWE |
| Testis, salivary glands, vessels | Characterising focal lesions; plaque and thrombus age | Strain, SWE |
| Intervention | Choosing the stiffest part of a lesion to biopsy; showing the ablation zone, which becomes stiff | Strain, ARFI |
Liver: diffuse liver disease
This is the best-validated use of elastography.
Why it works. Collagen deposited in fibrosis stiffens the liver. Stiffness rises with fibrosis stage, but not in a straight line: the steps are small between early stages and large between advanced fibrosis and cirrhosis. Elastography is therefore best at excluding or confirming advanced fibrosis and cirrhosis, and weaker at separating adjacent early stages.
Why it is wanted. Liver biopsy is invasive (pain, bleeding), samples about 1/50,000 of the liver, and suffers from sampling and observer variation. Fibrosis is graded on biopsy by METAVIR: F0 none, F1 portal fibrosis, F2 a few septa (significant fibrosis), F3 many septa (advanced fibrosis), F4 cirrhosis.
Technique for pSWE and 2D-SWE (SRU 2020 and EFSUMB 2017)
- Fasting: at least 4 hours (SRU); EFSUMB accepts a minimum of 2 hours, with 10 minutes of rest first.
- Supine or slight left lateral position, right arm raised above the head.
- Right intercostal approach, right lobe (segments 7 and 8); the left lobe gives higher and more variable values.
- Breath held in a neutral position: no deep inspiration and no Valsalva.
- Probe perpendicular to the capsule. Box 1.5-2 cm below the capsule, about 4-5 cm from the skin, clear of vessels, ducts and rib shadow.
- pSWE: 10 readings (at least 5 if IQR/M is within range). 2D-SWE: 5 readings when the machine has a quality map, 10 when it has none (SRU); EFSUMB accepts a minimum of 3. Report the median.
- Reliability: IQR/M of 30% or less for kPa, 15% or less for m/s (SRU).
- The report should state the machine and probe, the number of readings, median, IQR/M, and an interpretation that takes the cause of liver disease and the liver enzymes into account.
Cut-offs. Quote the source with every figure.
| Whose figures | Technique | Thresholds |
|---|---|---|
| SRU consensus 2020, "rule of four" (viral hepatitis and fatty liver disease) | pSWE and 2D-SWE, any vendor | 5 kPa (1.3 m/s) or less: normal. Under 9 kPa (1.7 m/s): compensated advanced chronic liver disease (cACLD) ruled out if there are no other clinical signs. 9-13 kPa (1.7-2.1 m/s): suggestive of cACLD, needs another test. Over 13 kPa (2.1 m/s): cACLD likely. Over 17 kPa (2.4 m/s): clinically significant portal hypertension likely |
| Baveno VII (2021), "rule of five" | Transient elastography | Under 10 kPa: cACLD excluded. 10-15 kPa: suggestive. Over 15 kPa: highly suggestive. 25 kPa or more: rules in clinically significant portal hypertension. 15 kPa or less with platelets 150 × 10⁹/L or more: rules it out. Under 20 kPa with platelets over 150: screening endoscopy for varices can be avoided |
| Castera 2005, chronic hepatitis C | Transient elastography | F2 or more: 7.1 kPa. F3 or more: 9.5 kPa. F4: 12.5 kPa |
| Friedrich-Rust 2012 meta-analysis | pSWE (Siemens VTQ) | F2 or more: 1.34 m/s. F3 or more: 1.55 m/s. F4: 1.80 m/s |
A normal liver measures about 4-6 kPa on transient elastography. Thresholds differ with the cause of liver disease (they tend to be higher in alcohol-related and cholestatic disease), so a reading is always interpreted against the aetiology. cACLD is the term that now replaces "advanced fibrosis and compensated cirrhosis" in non-invasive work, because elastography cannot separate F3 from F4 reliably.
What it is used for
- Staging fibrosis in chronic hepatitis B and C, metabolic dysfunction-associated steatotic liver disease (MASLD, formerly NAFLD), alcohol-related liver disease, autoimmune and cholestatic disease, and so deciding who needs treatment and who needs biopsy.
- Ruling cirrhosis in or out, which decides who enters surveillance for hepatocellular carcinoma and varices.
- Portal hypertension: liver stiffness correlates with the hepatic venous pressure gradient up to about 10-12 mmHg. Spleen stiffness adds to it: on transient elastography, under 21 kPa argues against and over 50 kPa argues for clinically significant portal hypertension, and 40 kPa or less makes high-risk varices unlikely (Baveno VII, validated mainly in viral hepatitis).
- Prognosis: higher stiffness predicts decompensation, hepatocellular carcinoma and death.
- Monitoring: response to antiviral treatment, abstinence or weight loss; progression on serial studies. Use the same machine each time.
- Steatosis: CAP on FibroScan, and attenuation imaging on newer scanners, quantify fat at the same sitting.
- Special groups: congestive hepatopathy and the Fontan circulation, Budd-Chiari syndrome, biliary atresia, cystic fibrosis, and graft fibrosis after transplantation.
- Focal liver lesions: malignant lesions are on average stiffer than benign ones, but the overlap is large; elastography is not recommended for characterising them.
Role of strain elastography in the liver: small, because there is no normal reference tissue in diffuse disease. A liver fibrosis index derived from strain patterns exists on some machines but is not in routine use.
Confounders that raise liver stiffness without fibrosis (false positives)
- Acute hepatitis and any flare with transaminases more than 5 times the upper limit of normal
- Obstructive cholestasis
- Hepatic congestion: right heart failure, outflow obstruction
- Recent meal (stiffness stays raised for 2-3 hours) and recent strenuous exercise
- Recent alcohol excess and acute alcoholic hepatitis
- Infiltration: amyloid, lymphoma, extramedullary haemopoiesis
- Deep inspiration or Valsalva; excess probe pressure on the left lobe
- Whether steatosis by itself changes the reading is debated
Causes of failed or unreliable readings: obesity and a thick chest wall, narrow intercostal spaces, ascites (transient elastography), inability to hold the breath, a small shrunken right lobe, and an inexperienced operator.
Ultrasound versus MR elastography for liver fibrosis
The details of MR elastography are in its own note; only what the comparison needs is given here.
MR elastography in four lines. A passive pneumatic driver strapped over the right lobe delivers continuous 60 Hz vibration from an active driver outside the scan room. A phase-contrast sequence with motion-encoding gradients synchronised to the vibration images the shear waves. An inversion algorithm converts the wave images into a stiffness map (elastogram) with a confidence map. Regions of interest are drawn over reliable liver and the mean is reported in kPa.
| Ultrasound elastography | MR elastography | |
|---|---|---|
| Liver sampled | Small: a 1 × 4 cm cylinder (TE) or a box of about 1 cm (pSWE); a few cm (2D-SWE) | Large: several slices through the whole liver, so least sampling error and best for patchy fibrosis |
| Accuracy | Very good for cirrhosis, moderate for early fibrosis | Highest of the non-invasive tests, particularly for early (F1-F2) fibrosis |
| Obesity and ascites | Fail or become unreliable (2D-SWE and pSWE cope with ascites; TE does not) | Not affected |
| Operator dependence | Moderate; needs training and quality criteria | Low; highly reproducible between readers and scanners |
| Failure | TE fails in about 3% and is unreliable in a further 15% or so, mostly from obesity | A few percent at 1.5 T, mostly from iron overload (signal loss on gradient-echo sequences); more at 3 T |
| Cost and access | Cheap, quick, portable, bedside, easily repeated | Expensive, limited availability, needs extra hardware and software |
| Time | 5-10 minutes | A few breath-holds, usually added to a liver MRI |
| Contraindications | None | Those of MRI: some implants, claustrophobia |
| Added information | B-mode and Doppler of liver and spleen; CAP for fat | Morphology, lesion detection and characterisation, fat fraction and iron in the same study |
| Standardisation | Values differ between vendors and techniques | One 2D gradient-echo technique at 60 Hz is used across vendors, so values are comparable |
| Confounders | Inflammation, congestion, cholestasis, food | The same; these are biological, not technical |
| Number reported | Young's modulus (or m/s) | Shear stiffness (magnitude of the complex shear modulus) |
The kPa values are not interchangeable. MR elastography reports shear stiffness, roughly a third of the Young's modulus reported by ultrasound, and it is measured at a different frequency. A normal liver is under about 2.5 kPa on MR elastography and cirrhosis is above about 5 kPa on the commonly used scale, against roughly 5 kPa and 13 kPa or more on ultrasound.
Bottom line: ultrasound elastography is the first-line screening and follow-up test; MR elastography is the problem-solver when ultrasound fails or is unreliable (obesity, ascites), when results disagree with the clinical picture, or when MRI is being done anyway.
Breast
Basis. Most cancers are harder than normal breast and harder than most benign lesions, because of the desmoplastic stromal reaction, which extends beyond the visible mass. Elastography is an adjunct to B-mode and BI-RADS, not a stand-alone test. Its main value is raising specificity and so reducing benign biopsies.
Technique. Patient positioned as for B-mode, linear probe perpendicular to the skin with the lightest possible contact, lesion in the centre of the box with fat and normal tissue around it, patient still or holding the breath. Excess pre-compression is the commonest error with either method.
Strain elastography: three ways to read it
- Tsukuba elasticity score (Itoh 2006), on a map where blue is hard
| Score | Pattern | Meaning |
|---|---|---|
| 1 | Whole lesion deforms like the surrounding tissue (all green) | Benign |
| 2 | Mosaic of green and blue | Probably benign |
| 3 | Centre blue, periphery green | Probably benign |
| 4 | Whole lesion blue | Suspicious |
| 5 | Lesion and surrounding tissue blue | Malignant |
The cut-off lies between 3 and 4. In the original series this gave a sensitivity of about 87% and a specificity of about 90%.
-
Strain ratio (fat-to-lesion ratio). One region of interest in the lesion, one in subcutaneous fat at a similar depth. Higher means stiffer. Published cut-offs range from about 2.3 to 4.5 depending on the machine and the study, so no single figure applies everywhere.
-
E/B ratio. Lesion length on the elastogram divided by its length on B-mode. Less than 1: benign. 1 or more: malignant. It reflects the stiff halo of desmoplasia and tumour infiltration that B-mode does not show.
Cysts on strain imaging
- Blue-green-red (BGR) sign: three horizontal colour layers inside a simple cyst on some machines.
- Bull's-eye artefact on others: a black lesion with a bright centre and a bright spot behind it.
- Both are artefacts of fluid moving under compression. They help to recognise a complicated cyst that looks solid on B-mode.
Shear wave elastography
- Read from the colour map (homogeneity, shape, and stiffness of the lesion and of the tissue around it) and from the number. Maximum stiffness (Emax), measured over the stiffest part of the lesion or its rim, performs best.
- BE1 multicentre study (Berg 2012, SuperSonic Imagine): a cut-off of 80 kPa (5.2 m/s). Adding shear wave findings to BI-RADS raised specificity from about 61% to about 78% without loss of sensitivity.
- Stiff rim sign: a ring of high stiffness around the lesion, typical of cancer. Very hard cancers may show no signal or falsely low values in the centre because shear waves do not propagate there, so the rim must be measured.
- Cysts show no signal (black), since shear waves do not travel in fluid.
Use with BI-RADS
- The ACR BI-RADS atlas, 5th edition (2013), lists elasticity assessment among the associated features of the ultrasound lexicon, with three descriptors: soft, intermediate, hard. It warns that shape, margin and echo pattern outweigh elasticity.
- Elastography changes management mainly in BI-RADS 3 and 4a:
- A BI-RADS 4a lesion that is soft on elastography (Tsukuba 1-3, or Emax of 80 kPa or less) can be downgraded to 3 and followed instead of biopsied.
- A BI-RADS 3 lesion that is hard (Tsukuba 4-5, or Emax of 160 kPa (7.3 m/s) or more in BE1) should be upgraded to biopsy.
- BI-RADS 4b, 4c and 5 lesions are biopsied whatever the elastogram shows. A soft result never cancels suspicious morphology.
Other roles in the breast
- Complicated cyst versus solid mass; fat lobule versus true lesion
- Showing the true extent of a cancer, which is often larger on the elastogram
- Choosing the stiffest part for biopsy
- Axillary nodes: metastatic nodes are stiffer, especially in the cortex
- Predicting and monitoring response to neoadjuvant chemotherapy (stiffness falls in responders)
- Non-mass lesions and architectural distortion, as a pointer to where to biopsy
Pitfalls
| False negative (soft cancers) | False positive (hard benign lesions) |
|---|---|
| Mucinous, medullary and papillary carcinoma | Hyalinised or calcified fibroadenoma |
| Necrotic or cystic high-grade tumours | Fat necrosis, surgical scar, radial scar |
| Ductal carcinoma in situ and very small invasive cancers | Mastitis, abscess wall, diabetic mastopathy |
| Lymphoma, some metastases | Sclerosing adenosis, fibrosis |
Technical limits: lesions deeper than about 4 cm, very large lesions that fill the box, lesions close to the chest wall or nipple, implants, and pre-compression.
Thyroid
Basis. Papillary carcinoma, the commonest thyroid cancer, is hard (fibrosis, psammoma bodies). Most benign nodules are soft. Elastography is added to greyscale ultrasound to decide which nodules need fine-needle aspiration.
Strain elastography
- Transverse or longitudinal plane, light compression. Carotid pulsation can be used as the compressing force, but it also causes artefact, so a longitudinal plane away from the artery is preferred.
- Rago score (5-point): 1, whole nodule elastic; 2, mostly elastic; 3, elastic at the periphery only; 4, whole nodule hard; 5, no elasticity in the nodule or in the area behind it. Scores 4 and 5 are suspicious.
- Asteria score (4-point): 1, entirely elastic; 2, mostly elastic; 3, mostly hard; 4, entirely hard. Scores 3 and 4 are suspicious.
- Strain ratio: normal thyroid (or strap muscle) strain divided by nodule strain. Reported cut-offs lie roughly between 2 and 4 and depend on the machine.
Shear wave elastography
- Gives kPa or m/s without manual compression, so it is less operator dependent.
- Normal thyroid is soft, roughly 10-20 kPa. Reported cut-offs for malignancy vary widely, from about 35 to 65 kPa or more for 2D-SWE and about 2.5-3 m/s for pSWE. No single cut-off is agreed.
Performance. Pooled sensitivity and specificity are each about 80% for both methods. The most useful property is a high negative predictive value: a nodule that is soft and has no suspicious greyscale feature is very unlikely to be malignant.
Role
- Adjunct to greyscale features in selecting nodules for aspiration, and in choosing which nodule to sample in a multinodular gland
- May help in nodules with indeterminate or non-diagnostic cytology, though results are mixed
- Follow-up of nodules with benign cytology
- Diffuse disease: stiffness is raised in Hashimoto, Graves, subacute (de Quervain) and Riedel thyroiditis
- Cervical lymph nodes: metastatic nodes are stiffer
Position in guidelines. WFUMB (2017) and EFSUMB recommend it only as a complement to greyscale ultrasound, never a replacement. It is not a scoring item in ACR TI-RADS (2017); EU-TIRADS mentions it as an optional complementary tool; the earlier French TI-RADS included high stiffness as a suspicious sign.
Limitations
- Follicular carcinoma is often soft and is missed; medullary and some other cancers can be soft too
- Hard benign nodules: coarse or rim calcification, fibrosis, subacute thyroiditis
- Cystic or predominantly cystic nodules are unreliable (fluid)
- Nodules larger than about 3 cm, with no normal tissue left for reference; coalescent nodules
- Isthmus (compressed against the trachea), deep or retrosternal nodules
- Background thyroiditis makes the reference tissue stiff
- Carotid pulsation, swallowing, breathing; pre-compression raises all values
Obstetrics and gynaecology
Most of these uses are still under evaluation; say so in the answer.
Obstetrics
| Use | What is found |
|---|---|
| Cervix: prediction of preterm birth | The cervix softens (collagen remodelling) before it shortens. A soft cervix, especially at the internal os, in the mid-trimester is associated with spontaneous preterm delivery, and adds to cervical length |
| Cervix: before induction of labour | A soft cervix predicts successful induction and a shorter induction-to-delivery time; proposed as an objective addition to the Bishop score |
| Cervical insufficiency and cerclage | Assessment of cervical consistency before and after cerclage |
| Placenta | Stiffer in pre-eclampsia, fetal growth restriction and gestational diabetes in research series |
| Placenta accreta spectrum, caesarean scar | Investigational |
| Fetal organs (lung maturity, liver) | Research only |
| Post-partum | Pelvic floor: levator ani and anal sphincter injury |
- Technique for the cervix: transvaginal probe in the anterior fornix, sagittal view as for cervical length, minimal pressure. Strain imaging is read as a colour pattern or as strain values at the internal and external os; the weakness is that there is no reference tissue and the applied force is unknown. Shear wave imaging gives an absolute value but is affected by depth and by probe pressure.
- Safety: strain elastography uses the same output as B-mode. ARFI-based shear wave methods use longer, higher-energy pulses and raise the thermal index, so they should be used on the cervix and placenta with the ALARA principle, kept away from the fetus, and are not recommended for routine fetal imaging.
Gynaecology
| Use | What is found |
|---|---|
| Fibroid versus adenomyosis | A fibroid is a well-defined, usually stiff lesion with a sharp border and a soft rim (pseudocapsule). Adenomyosis is ill-defined and heterogeneous, with no border. Both are generally stiffer than normal myometrium; which of the two is stiffer differs between studies, so the pattern matters more than the number |
| Fibroid treatment | Mapping before, and showing the stiff treated zone after, high-intensity focused ultrasound or embolisation |
| Endometrium | Carcinoma is stiffer than polyp or hyperplasia; may help in assessing myometrial invasion |
| Cervical cancer | Stiffer than normal cervix; helps to define tumour extent and parametrial spread and to monitor chemoradiation |
| Ovary and adnexa | Solid parts of malignant masses are stiffer; overlap is large and it does not replace the standard ultrasound rules. Endometriomas and deep endometriotic nodules are stiff |
| Pelvic floor | Levator ani stiffness in prolapse and incontinence |
| Breast in pregnancy and lactation | As for breast generally |
Limitations in this field: no agreed cut-offs, different machines and methods across studies, deep pelvic organs beyond the reach of transabdominal elastography, bowel and bladder motion, and unresolved fetal safety for ARFI.
Clinical applications of compression elastography
Compression (strain) elastography shows relative stiffness, so its uses are in focal lesions surrounded by normal tissue in superficial organs or organs reached by an endocavitary probe. Write the technique and the three ways of reading it from the strain elastography section, then the following.
| Organ | Application | How it is read |
|---|---|---|
| Breast (the main use) | Benign versus malignant; downgrading BI-RADS 4a and upgrading BI-RADS 3; complicated cyst versus solid; lesion extent; biopsy targeting; axillary nodes | Tsukuba score (cut-off between 3 and 4), strain ratio against fat, E/B ratio of 1 or more, BGR sign in cysts |
| Thyroid | Selecting nodules for aspiration; high negative predictive value | Rago score 4-5 or Asteria 3-4 suspicious; strain ratio against normal gland |
| Lymph nodes | Metastatic versus reactive; which node to sample | Proportion of hard area; strain ratio against muscle |
| Prostate (transrectal) | Targeting biopsy of hard peripheral-zone foci; fewer cores | Reproducible hard focus on repeated compression |
| Cervix (transvaginal) | Preterm birth risk, induction success, cervical cancer extent | Colour pattern, strain at internal os |
| Uterus | Fibroid versus adenomyosis; endometrial lesions | Border and pattern of the stiff area |
| Musculoskeletal | Achilles, patellar and common extensor tendinopathy; plantar fasciitis; muscle tear; soft tissue masses | Softening of a normally hard tendon; colour grade |
| Endoscopic ultrasound | Pancreatic mass (cancer versus chronic pancreatitis); mediastinal and abdominal nodes; rectal wall | Dominant colour, strain ratio, strain histogram |
| Testis and scrotum | Hard tumour versus soft benign lesion or haematoma | Colour score |
| Salivary glands, skin and subcutis | Characterising masses; scleroderma | Colour score, ratio |
| Liver | Limited: fibrosis index from strain pattern; showing the stiff ablation zone after radiofrequency ablation | Pattern analysis |
| Vessels | Plaque composition; age of venous thrombus (older thrombus is harder) | Strain within plaque or thrombus |
End with its limits: relative, operator dependent, shallow, unsuitable for diffuse disease, and not comparable between vendors.
Advantages
- Non-invasive, quick, no radiation, no contrast, bedside, repeatable
- Added to a routine ultrasound examination on the same machine and probe
- Shear wave methods give a number that can be followed over time
- Reduces liver biopsies and benign breast and thyroid biopsies
- Cheap compared with MR elastography
Limitations, artefacts and pitfalls
General
- Values and cut-offs are specific to technique, vendor, probe and disease; follow-up should be on the same system
- Depth: strain to 3-4 cm, ARFI to about 6-7 cm
- Assumes simple elastic behaviour; tissue is viscoelastic, anisotropic (muscle, tendon, kidney) and stiffens under pressure
- Operator training matters, most of all for strain imaging
- Stiffness is not specific: fibrosis, tumour, inflammation, congestion and oedema all raise it
Artefacts
| Artefact | Cause |
|---|---|
| Pre-compression artefact | Probe pressure stiffens superficial tissue and raises all values |
| BGR and bull's-eye artefacts | Fluid in a cyst on strain imaging |
| Signal void | No shear wave in fluid, or in very hard or deep tissue; shown as black or uncoloured |
| "Bang" or near-field artefact | High values just below the liver capsule from reverberation; hence the 1.5-2 cm rule |
| Stiff rim, soft centre | Shear waves fail to propagate through the centre of a hard cancer |
| Slip and out-of-plane motion | Lesion moves sideways under compression |
| Shadowing | Calcification or ribs block the push and the tracking pulses |
| Motion | Breathing, cardiac and carotid pulsation, swallowing |
| Boundary effects | Shear waves reflect at capsules, vessels and thin structures and give falsely high or low values |
Safety. ARFI pulses are longer than imaging pulses but remain within regulatory output limits; the main caution is the fetus.
Short note (5 marks)
- Definition: imaging of tissue stiffness, the equivalent of palpation. Young's modulus E = stress / strain, in kPa.
- Strain (compression) elastography: manual or physiological compression; strain is tracked and shown as a colour map. Relative only: score, strain ratio, E/B ratio.
- Shear wave elastography: a piston (transient elastography, 50 Hz) or an ARFI push (point and 2D shear wave) produces shear waves; their speed (1-10 m/s) gives E = 3ρc². Quantitative, in m/s or kPa. Small diagram.
- Liver: fibrosis staging and portal hypertension. Ten readings, median, IQR/M of 30% or less, after fasting. SRU 2020 rule of four for ARFI: 5, 9, 13, 17 kPa. Baveno VII for transient elastography: under 10 kPa excludes and over 15 kPa suggests advanced chronic liver disease.
- Confounders: inflammation with high transaminases, cholestasis, congestion, a recent meal.
- Breast: Tsukuba score 1-5 (cut-off between 3 and 4), E/B ratio of 1 or more, Emax above 80 kPa suspicious; refines BI-RADS 3 and 4a.
- Thyroid and others: adjunct for nodule selection (follicular carcinoma is soft); cervix for preterm birth; prostate, nodes, tendon.
- Limits: vendor-specific values, depth, operator dependence, obesity and ascites; MR elastography samples more liver and is used when ultrasound fails.